Prosthetic Kit

iRES® KIT

Prosthetic Kit

The purpose of surgical trays is to store the instruments used to insert dental implants. The kit can be carried, sterilized and kept in a horizontal position with the lid closed. All the instruments must be cleaned and sterilized before the first use. The surgical kit and instruments are not sterile at the time of delivery.

Discover our promotions implants + kit

Prosthetic kit

#1 Cleaning

1] Dismantle all the compound parts.
2] Rinse abundantly with cold or lukewarm water for 2-5 minutes.

3] Leave the instruments for 10 minutes in an ultrasonic cleaner with a neutral pH. enzymatic detergent diluted in water according to the product instructions. 4] Wash the instruments with water for 3 minutes.


Sterilization

The guidelines for sterilization are listed below. Exceeding these sterilization limits may cause deterioration of the plastic components.

type of cycle (value)
Temperature (°C/F)
Exposure
Drying time
Pre-vacuum
132 / 270
3 minutes
30 minutes
Pre-vacuum
134 / 273
18 minutes
30 minutes
Gravity
121 / 250
80 minutes
30 minutes

Prosthetic kit composition

Code
Description
MDS
short screwdriver hex 1.25 mm h 23 mm
MDL
long screwdriver hex 1.25 mm h 29 mm
TRT
removal tool for abutments
HDH20
implant driver for straight MUA
THDDS
short prosthetic screwdriver hex 1.25
THDDL
long prosthetic screwdriver hex 1.25
TWA2
ratchet wrench
MDLA (on request)
angled screwdriver hex 1.25 mm h 29 mm