iRES® KIT
The purpose of surgical trays is to store the instruments used to insert dental implants. The kit can be carried, sterilized and kept in a horizontal position with the lid closed. All the instruments must be cleaned and sterilized before the first use. The surgical kit and instruments are not sterile at the time of delivery.
Discover our promotions implants + kitProsthetic kit
1] Dismantle all the compound parts.
2] Rinse abundantly with cold or lukewarm water for 2-5 minutes.
3] Leave the instruments for 10 minutes in an ultrasonic cleaner with a neutral pH. enzymatic detergent diluted in water according to the product instructions.
4] Wash the instruments with water for 3 minutes.
The guidelines for sterilization are listed below. Exceeding these sterilization limits may cause deterioration of the plastic components.
type of cycle (value) | Temperature (°C/F) | Exposure | Drying time |
---|---|---|---|
Pre-vacuum | 132 / 270 | 3 minutes | 30 minutes |
Pre-vacuum | 134 / 273 | 18 minutes | 30 minutes |
Gravity | 121 / 250 | 80 minutes | 30 minutes |
Code | Description |
---|---|
MDS | short screwdriver hex 1.25 mm h 23 mm |
MDL | long screwdriver hex 1.25 mm h 29 mm |
TRT | removal tool for abutments |
HDH20 | implant driver for straight MUA |
THDDS | short prosthetic screwdriver hex 1.25 |
THDDL | long prosthetic screwdriver hex 1.25 |
TWA2 | ratchet wrench |
MDLA (on request) | angled screwdriver hex 1.25 mm h 29 mm |